Entry time: Fri Oct 24 11:42:48 2008
|
Author: |
Alberto |
Type: |
Update |
Category: |
General |
Subject: |
Abs length: locking the X arm cavity in TEM01/10 |
|
|
I went back to lock the arm cavity in either TEM01 or TEM10 mode. Attached are the results. We still have several resonances which we can't clearly identify. I expect TEM01/10 to be at 6.276MHz but we don't have a peak exactly there. What we have is:
- a peak at 6.320MHz in the measurement of the TEM01 mode (the one with the lobes of the spot almost on the vertical axis)
- a peak at 6.590MHz in both the TEM01 and TEM10 measurements.
I'm either missing the real TEM01/10 mode or the peaks at 6.590MHz are those. If that were true, that would mean that the radius of curvature of ETMX is 49.29 m instead of 57.57 m as listed in the IFO data sheets. I think it's much more likely that the measurements are missing the right peaks. |
|
|