Entry  Mon Oct 6 10:04:58 2014, manasa, Update, General, Unexpected power shutdown 
    Reply  Mon Oct 6 10:23:43 2014, Steve, Update, VAC, Unexpected power shutdown 
       Reply  Tue Oct 7 16:55:16 2014, Steve, Update, VAC, Unexpected sweaty valves sweatyV5.jpgsweatyV5f.jpg
          Reply  Wed Oct 8 08:46:57 2014, Steve, Update, VAC, VAT valves actuator lubricant 
             Reply  Mon Oct 13 14:44:52 2014, Steve, Update, VAC, RGA scan pd78 -day 13 RGAscan13Oct2014.png
                Reply  Mon Dec 1 17:13:57 2014, Steve, Update, VAC, RGA scan pd78 -day 63 RGAscanD63.png
                   Reply  Tue Mar 31 09:03:32 2015, Steve, Update, VAC, RGA scan pd78 day 182 RGApd78d182.png
                      Reply  Wed Jul 1 10:01:25 2015, Steve, Update, VAC, RGA scan pd78 day 275 rga275d.png
                         Reply  Mon Oct 5 10:11:35 2015, Steve, Update, VAC, RGA scan pd78 day 386 RGA_background.png
       Reply  Tue Dec 23 14:33:24 2014, Steve, Update, VAC, Chiara gets UPS prepUP.png
    Reply  Mon Oct 6 10:28:18 2014, manasa, Update, General, Unexpected power shutdown 
       Reply  Mon Oct 6 11:09:52 2014, Jenne, Update, General, Unexpected power shutdown: slow computers 
       Reply  Mon Oct 6 17:04:51 2014, ericq, Update, General, Unexpected power shutdown 
          Reply  Mon Oct 6 17:36:17 2014, ericq, Update, General, Unexpected power shutdown 
          Reply  Mon Oct 6 18:15:12 2014, manasa, Update, General, Unexpected power shutdown: end green alignment 
    Reply  Tue Oct 7 16:17:15 2014, Steve, Update, General, reason for 8 sec power outage 
Message ID: 10837     Entry time: Tue Dec 23 14:33:24 2014     In reply to: 10568
Author: Steve 
Type: Update 
Category: VAC 
Subject: Chiara gets UPS 

Quote:

Quote:

We had an unexpected power shutdown for 5 sec at ~ 9:15 AM.

Chiara had to be powered up and am in the process of getting everything else back up again.

Steve checked the vacuum and everything looks fine with the vacuum system.

PSL Innolight laser and the 3 units of IFO air conditions turned on.

The vacuum system reaction to losing power: V1 closed and Maglev shut down. Maglev is running on 220VAC so it is not connected to VAC-UPS.  V1 interlock was triggered by Maglev "failure" message.

Maglev was reset and started. After Chiara was turned on manually I could bring up the vac control screen through Nodus and opened V1

"Vacuum Normal" valve configuration was recovered instantly.

 

Chiara needs UPS 

It is arriving Thursday

 EricQ and Steve,

Steve preset the vacuum for safe-reboot mode with C1vac1 and C1vac2 running normal: closed valves as shown, stopped Maglev & disconnected valves V1 plus valves with moving labels.

(The position indicator of the valves changes to " moving " when its cable disconnected )

Eric shut down Chiara, installed APC's UPS Pro 1000 and restarted it.

All went well. Nothing unexpected happened. So we can conclude that the vacuum system with running C1vac1 and C1vac2 is not effected by Chiara's losing AC power.

Attachment 1: prepUP.png  27 kB  | Hide | Hide all
prepUP.png