Entry  Wed Nov 12 18:08:35 2014, Jenne, Update, LSC, RIN in transmission a problem? RIN_TRX_TRY_MCtrans_12Nov2014_zip.xml.gzRIN_TRX_TRY_MCtrans_12Nov2014.pdfRINcontribution_12Nov2014_linearXscale.pngRINcontribution_12NOv2014_logXscale.png
    Reply  Thu Nov 13 01:00:37 2014, rana, Update, LSC, RIN in transmission a problem? 
    Reply  Thu Nov 13 04:28:28 2014, Jenne, Update, LSC, RIN in transmission a problem? zippyzip.zipRIN_TRX_TRY_IPPOS_NoLens_12Nov2014.pdf
       Reply  Thu Nov 13 23:42:01 2014, Jenne, Update, LSC, RIN vs. Seismic Seismic_TRXTRYandMC_13Nov2014.pdf
          Reply  Fri Nov 14 17:08:17 2014, Jenne, Update, LSC, RIN vs. Seismic 
       Reply  Fri Nov 14 20:31:13 2014, ericq, Update, LSC, RIN in transmission a problem? 
Message ID: 10712     Entry time: Thu Nov 13 23:42:01 2014     In reply to: 10709     Reply to this: 10718
Author: Jenne 
Type: Update 
Category: LSC 
Subject: RIN vs. Seismic 

After Kate, Diego and I moved the seismometers to the corner for a huddle test (see elog 10711), I wanted to check the coherence between the seismometers and the arm transmissions.  

First of all, it looks like either the Guralp or the T-240 have their X and Y backwards. Diego is going to check this tomorrow. 

Between 0.9Hz - 3.5Hz, we have pretty strong coherence with the horizontal seismic channels.  Interestingly, between 8-10Hz, the Yarm has pretty strong coherence with the Z-axes of the seismometers (the coherence is only about 0.6, but it's consistent over a 2-ish Hz wide band). 

The MC transmission doesn't have as much coherence with the seismic, which surprised me.  So, we can try some FF to the MC, but we may also have to do some to the arms.

Seismic_TRXTRYandMC_13Nov2014.pdf