Wed Nov 5 01:58:54 2014, ericq, Update, LSC, 3F RFPD RF spectra  
|
Wed Nov 5 11:09:44 2014, Koji, Update, LSC, 3F RFPD RF spectra
|
Wed Nov 5 22:25:42 2014, ericq, Update, LSC, 3F RFPD RF spectra
|
Thu Nov 6 01:58:55 2014, Koji, Update, LSC, 3F RFPD RF spectra
|
Thu Nov 6 11:49:58 2014, ericq, Update, LSC, 3F RFPD RF spectra
|
Thu Nov 6 14:41:49 2014, Koij, Update, LSC, 3F RFPD RF spectra
|
Fri Nov 7 02:21:12 2014, ericq, Update, LSC, 3F RFPD RF spectra
|
Fri Nov 7 14:41:18 2014, ericq, Update, LSC, 3F RFPD RF spectra 
|
Fri Nov 7 16:15:53 2014, Jenne, Update, LSC, 3F RFPD RF spectra
|
Sat Nov 8 11:35:05 2014, rana, Update, LSC, 3F RFPD RF spectra
|
Mon Nov 10 18:11:57 2014, ericq, Update, LSC, 3F RFPD RF spectra
|
|
Message ID: 10679
Entry time: Thu Nov 6 11:49:58 2014
In reply to: 10675
Reply to this: 10682
|
Author: |
ericq |
Type: |
Update |
Category: |
LSC |
Subject: |
3F RFPD RF spectra |
|
|
Quote: |
Where is the PD out spectrum measured with the coupler???
|
The "coupled" port of the coupler went to the AG4395 input, the output of the Highpass is connected to the "IN", and the "OUT" goes to the demod board. |