Entry  Wed Nov 5 17:50:29 2014, manasa, Update, General, PSL and AUXY beatnote in IR found PSLsetup.jpg
    Reply  Wed Nov 5 18:08:00 2014, ericq, Update, LSC, PSL and AUXY beatnote in IR found 
Message ID: 10672     Entry time: Wed Nov 5 18:08:00 2014     In reply to: 10671
Author: ericq 
Type: Update 
Category: LSC 
Subject: PSL and AUXY beatnote in IR found 

Green beatnotes recovered.

It was just a matter of aligning the arm greens and PSL greens on the PSL table. I suppose something knocked the PSL alignment out of whack... I was also able to simultaneously see the green beatnote and IR beatnote respond to Yend laser temperature. 

Locked arms on POX/POY, checked RMS of ALS-BEAT[X/Y]_FINE_PHASE_OUT_HZ channels. 

  • ALSY: 300Hz RMS
  • ALSX: 700Hz RMS

These seem fine. Locked CARM and DARM on ALS, found IR resonances. 

ALS is back in business 