40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Mon Oct 6 10:04:58 2014, manasa, Update, General, Unexpected power shutdown 
    Reply  Mon Oct 6 10:23:43 2014, Steve, Update, VAC, Unexpected power shutdown 
       Reply  Tue Oct 7 16:55:16 2014, Steve, Update, VAC, Unexpected sweaty valves sweatyV5.jpgsweatyV5f.jpg
          Reply  Wed Oct 8 08:46:57 2014, Steve, Update, VAC, VAT valves actuator lubricant 
             Reply  Mon Oct 13 14:44:52 2014, Steve, Update, VAC, RGA scan pd78 -day 13 RGAscan13Oct2014.png
                Reply  Mon Dec 1 17:13:57 2014, Steve, Update, VAC, RGA scan pd78 -day 63 RGAscanD63.png
                   Reply  Tue Mar 31 09:03:32 2015, Steve, Update, VAC, RGA scan pd78 day 182 RGApd78d182.png
                      Reply  Wed Jul 1 10:01:25 2015, Steve, Update, VAC, RGA scan pd78 day 275 rga275d.png
                         Reply  Mon Oct 5 10:11:35 2015, Steve, Update, VAC, RGA scan pd78 day 386 RGA_background.png
       Reply  Tue Dec 23 14:33:24 2014, Steve, Update, VAC, Chiara gets UPS prepUP.png
    Reply  Mon Oct 6 10:28:18 2014, manasa, Update, General, Unexpected power shutdown 
       Reply  Mon Oct 6 11:09:52 2014, Jenne, Update, General, Unexpected power shutdown: slow computers 
       Reply  Mon Oct 6 17:04:51 2014, ericq, Update, General, Unexpected power shutdown 
          Reply  Mon Oct 6 17:36:17 2014, ericq, Update, General, Unexpected power shutdown 
          Reply  Mon Oct 6 18:15:12 2014, manasa, Update, General, Unexpected power shutdown: end green alignment 
    Reply  Tue Oct 7 16:17:15 2014, Steve, Update, General, reason for 8 sec power outage 
Message ID: 10579     Entry time: Tue Oct 7 16:55:16 2014     In reply to: 10568     Reply to this: 10584
Author: Steve 
Type: Update 
Category: VAC 
Subject: Unexpected sweaty valves 

 Pump  spool valves V5, V4, V3 sweating a lot. VM3 and VC2 not so much.

They are VAT valves F28-62887-03, 11, 14 and so on ~15-16 years old.

 I'm speculating that some plastic is aging-braking down at the atmospheric-pneumatic side of valves.
The vacuum side is not effected, according to vacuum pressure readings.

May be some condensation from the small turbos? No

I'm looking for an identical valve to examine, but I can not find one.

We are using industrial grade 99.96% Nitrogen to actuate these valves.

Valves are not effected are  dry: VA6, V6, V7 and all annuloses.

 