Thu Sep 25 19:42:12 2014, ericq, Update, General, Vent progress
|
Fri Sep 26 12:25:34 2014, ericq, Update, General, Vent progress
|
Fri Sep 26 16:10:14 2014, ericq, Update, General, Vent update
|
Fri Sep 26 17:13:39 2014, ericq, Update, General, Vent update
|
Mon Sep 29 12:47:51 2014, ericq, Update, General, Vent update
|
Mon Sep 29 17:10:51 2014, ericq, Update, General, Vent update
|
Mon Sep 29 18:12:24 2014, ericq, Update, General, Vent update
|
Tue Sep 30 11:53:29 2014, ericq, Update, General, Vent update
|
Tue Sep 30 16:18:57 2014, Steve, Update, VAC, pump down #78 has started
|
Tue Sep 30 18:02:53 2014, Steve, Update, VAC, pump down #78 stops at 320 Torr
|
Wed Oct 1 10:21:19 2014, Steve, Update, VAC, pump down #78 stops at 3.5 Torr
|
Wed Oct 1 16:26:53 2014, Steve, Update, VAC, pump down #78 completed
|
Thu Oct 2 09:23:23 2014, Steve, Update, VAC, vent 77
|
Fri Oct 3 10:10:37 2014, Steve, Update, VAC, cold cathode gauge reading switched
|
Fri May 8 14:30:09 2015, Steve, Update, VAC, CC1 cold cathode gauges are baked now
|
Tue May 12 14:57:52 2015, Steve, Update, VAC, CC1 cold cathode gauges are baked now
|
|
Message ID: 10545
Entry time: Fri Sep 26 16:10:14 2014
In reply to: 10544
Reply to this: 10546
|
Author: |
ericq |
Type: |
Update |
Category: |
General |
Subject: |
Vent update |
|
|
Today so far:
- I moved SRM forward by 3mm
- Then I leveled the ITMY table
- At this point, bringing the ITMY oplev beam back onto its QPD got me back to green locking and IR flashes
- AS and POY beams are both making it out to their tables, as seen by IR card. (Though not to their in-air optics)
Here's my quick brain dump of things to do before we can pump down (anyone see anything missing?):
- Check the clearance of the POY beam at the SRM cage
- Re-do distance reconstruction measurements, confirm desired SRC length
- Lock the SRM cage down fully (right now, has 2 clamps on, and one laying unused)
- Align SRM for SRC flashes
- Adjust SRM OSEM positions as needed
- Adjust SRM oplev beam path, measure lever arm for calibration
- Confirm beam spots on output mirrors in ITMY and BS chambers are ok
- Take pictures of ITMY chamber.
- Closeup checklist
|