We made quantitative inspection of the X/Y green beat setup on the PSL table.
DC output of the BBPD for each arm was measured by blockiing the beams at either or both side of the recombination BS.
The power over lap for the X arm beat note setup was 7.8% and is now 53%.
There is 3dB of headroom for the improvement of the mode overlap.
The power over lap for the Y arm beat note setup was 1.2% and is now 35%.
There is 4dB of headroom for the improvement of the mode overlap.
The RF analyzer monitor for the beat power is about 10dB lower than expected. Can we explain this only by the cable loss?
If not it there something causing the big attenuation?
XARM YARM
o BBPD DC output (mV)
V_DARK: - 3.3 + 1.9
V_PSL: + 4.3 +22.5
V_ARM: +187.0 + 8.4
o BBPD DC photocurrent (uA)
I_DC = V_DC / R_DC ... R_DC: DC transimpedance (2kOhm)
I_PSL: 3.8 10.3
I_ARM: 95.0 3.3
o Expected beat note amplitude
I_beat_full = I1 + I2 + 2 sqrt(e I1 I2) cos(w t) ... e: mode overwrap (in power)
I_beat_RF = 2 sqrt(e I1 I2)
V_RF = 2 R sqrt(e I1 I2) ... R: RF transimpedance (2kOhm)
P_RF = V_RF^2/2/50 [Watt]
= 10 log10(V_RF^2/2/50*1000) [dBm]
= 10 log10(e I1 I2) + 82.0412 [dBm]
= 10 log10(e) +10 log10(I1 I2) + 82.0412 [dBm]
for e=1, the expected RF power at the PDs [dBm]
P_RF: -12.4 -22.6
o Measured beat note power (before the alignment)
P_RF: -23.5 -41.7 [dBm] (38.3MHz and 34.4MHz)
e: 7.8 1.2 [%]
o Measured beat note power (after the alignment)
P_RF: -15.2 -27.1 [dBm] (26.6MHz and 26.8MHz)
e: 53 35 [%]
Measured beat note power at the RF analyzer in the control room
P_CR: -25 -20 [dBm]
Expected -17 - 9 [dBm]
Expected Power:
Pin + External Amp Gain (0dB for X, 20dB for Y)
- Isolation trans (1dB)
+ GAV81 amp (10dB)
- Coupler (10.5dB)
|