40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Wed Aug 27 18:03:47 2014, ericq, Update, Green Locking, uPDH Box Checkup uPDH_X.JPGuPDH_Y.JPGupdhTFs.pdfupdhNoises.pdfuPDH_Aug27.zip
    Reply  Thu Aug 28 17:34:20 2014, ericq, Update, Green Locking, uPDH Box Checkup TEK00006.PNG
       Reply  Fri Aug 29 17:28:07 2014, rana, Update, Green Locking, uPDH Box Checkup 
          Reply  Fri Aug 29 23:53:35 2014, ericq, Update, Green Locking, uPDH Box Checkup 
             Reply  Tue Sep 2 20:02:06 2014, ericq, Update, Green Locking, uPDH Box Checkup 
       Reply  Wed Sep 3 00:17:22 2014, rana, Update, Green Locking, uPDH Box Checkup 
          Reply  Wed Sep 3 20:38:40 2014, ericq, Update, Green Locking, uPDH Box Checkup 
             Reply  Fri Sep 5 04:07:44 2014, ericq, Update, Green Locking, X end uPDH Box Replaced 
                Reply  Fri Sep 5 16:13:00 2014, ericq, Update, Green Locking, new X Green PDH measurements 6x
                   Reply  Sat Sep 6 01:48:54 2014, rana, Update, Green Locking, new X Green PDH measurements 
                      Reply  Sat Sep 6 14:49:12 2014, ericq, Update, Green Locking, new X Green PDH measurements xGComparison.pdfxGspectra.pdf
                         Reply  Sun Sep 7 17:06:38 2014, rana, Update, Green Locking, new X Green PDH measurements 
                         Reply  Mon Sep 8 14:14:13 2014, Jenne, Update, Green Locking, new Y Green PDH measurements 6x
Message ID: 10463     Entry time: Sat Sep 6 01:48:54 2014     In reply to: 10460     Reply to this: 10464
Author: rana 
Type: Update 
Category: Green Locking 
Subject: new X Green PDH measurements 

Quote:

Does this matter? Is there something in my simulation that I can correct that would give a more accurate calibration? 

Data, plots, code, attached. 

 What modulation depth are you using for the simulation? I have never seen a real measurement of that in our elog for the end-PDH systems.

I also disbelieve your RMS calculations. It looks like in the 1.5-0.5 Hz band we're picking up 50 kHz of frequency noise even though the 1 Hz peak is only 80 Hz/rHz, even though math says  "80 * sqrt(1) = 80".

Take a look at:

http://www.mathwords.com/r/root_mean_square.htm

and

http://www.ligo.caltech.edu/~rana/mat/utilities/rms.m