Message ID: 10450     Entry time: Thu Sep 4 03:12:55 2014     In reply to: 10448
Author: ericq 
Type: Update 
Category: LSC 
Subject: Green PDH box boosts 

Jenne made her board modifications, and the measured TF agreed with the design. Alas, the green would not lock to the arm in this state. 

I think that the reason is that the new TF does not have nearly as much low frequency gain as the old one, for a given UGF. Thus, for example, the 1Hz noise due to the pendulum resonance, has 30dB less loop gain suppressing it. 

boostedTF.pdf

 

NEED MORE gain.jpg

 