Entry  Wed Aug 27 18:03:47 2014, ericq, Update, Green Locking, uPDH Box Checkup uPDH_X.JPGuPDH_Y.JPGupdhTFs.pdfupdhNoises.pdfuPDH_Aug27.zip
    Reply  Thu Aug 28 17:34:20 2014, ericq, Update, Green Locking, uPDH Box Checkup TEK00006.PNG
       Reply  Fri Aug 29 17:28:07 2014, rana, Update, Green Locking, uPDH Box Checkup 
          Reply  Fri Aug 29 23:53:35 2014, ericq, Update, Green Locking, uPDH Box Checkup 
             Reply  Tue Sep 2 20:02:06 2014, ericq, Update, Green Locking, uPDH Box Checkup 
       Reply  Wed Sep 3 00:17:22 2014, rana, Update, Green Locking, uPDH Box Checkup 
          Reply  Wed Sep 3 20:38:40 2014, ericq, Update, Green Locking, uPDH Box Checkup 
             Reply  Fri Sep 5 04:07:44 2014, ericq, Update, Green Locking, X end uPDH Box Replaced 
                Reply  Fri Sep 5 16:13:00 2014, ericq, Update, Green Locking, new X Green PDH measurements 6x
                   Reply  Sat Sep 6 01:48:54 2014, rana, Update, Green Locking, new X Green PDH measurements 
                      Reply  Sat Sep 6 14:49:12 2014, ericq, Update, Green Locking, new X Green PDH measurements xGComparison.pdfxGspectra.pdf
                         Reply  Sun Sep 7 17:06:38 2014, rana, Update, Green Locking, new X Green PDH measurements 
                         Reply  Mon Sep 8 14:14:13 2014, Jenne, Update, Green Locking, new Y Green PDH measurements 6x
Message ID: 10441     Entry time: Tue Sep 2 20:02:06 2014     In reply to: 10439
Author: ericq 
Type: Update 
Category: Green Locking 
Subject: uPDH Box Checkup 

I narrowed down the saturation point in the X green PDH box to the preamp inside the AD8336, but there is still no clear answer as to why it's happening. 

As per Jenne's request, I put the X end PDH box back for tonight's work. It locks, but we have an artificially low actuation range. With SR785, I confirmed a PDH UGF around 5k. Higher than that, and I couldn't reliably measure the UGF due to SR560 saturations. The analyzer is not currently in the loop. 

Both arms lock to green, but I haven't looked at beatnotes today. 