Message ID: 10391     Entry time: Thu Aug 14 19:23:25 2014
Author: rana 
Type: HowTo 
Category: IOO 
Subject: How do I set the FSS offset to make the PZT voltage start at the right place? 

 When the IMC locks, we want the FAST OUT of the TTFSS box to be close to zero volts. We also want the control signal from the MC Servo board to be close to 0 V. How to set this up?

With the IMC locked, we just servo the FSS input offset to minimize the MC board output :

ezcaservo -r C1:IOO-MC_FAST_MON -g 0.1 -t 10 C1:PSL-FSS_INOFFSET

I would have used "CDSUTILS", but that seems to have some sort of ridiculous bug where we can't have prefixes on channel names, even on the command line. 