The PZT seems to saturate at around +/- 3500 counts. So for the Y arm, I excluded the saturated points and fitted the data points again.
As for the calibration number, we expect the 3276.8 count/V for +/- 10 V range of a 16 bit ADC but the number is ~800 count/V. I couldn't figure out a reason why the number is so different.
The new calibration values are :
X- arm PZT : [146.3 +/- 2.37 ] counts/Volt (with a 20 dB attenuator included in the path)
Y- arm PZT : [ 797 +/- 3.6] counts/Volt
I will get the calibration in MHz/V of PZT actuation and check whether these numbers make any sense. |