40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Thu Jul 17 17:08:58 2014, Harry, Update, General, 1X2 Rack Changes 1X2Changes.png
    Reply  Thu Jul 17 21:01:28 2014, Manasa, Update, General, 1X2 Rack Changes 
       Reply  Thu Jul 17 22:08:14 2014, Koji, Update, General, 1X2 Rack Changes 
Message ID: 10230     Entry time: Thu Jul 17 17:08:58 2014     Reply to this: 10233
Author: Harry 
Type: Update 
Category: General 
Subject: 1X2 Rack Changes 

 Purpose

 

Steve and I moved some things around in the 1X2 rack in order to make room (roughly 6") for the electronics box that will contain rf frequency counters, ADC, and Raspberry Pi's for use in the Frequency Offset Locking apparatus

Picture

1X2Changes.png

Occurrences

First, we killed power by removing the fuse that the boxes we were moving were running through.

Then, we moved the boxes. I dropped//lost a washer. It didn't seem to cause any problems, so no further attempts to locate it were made.

The fuse was reinstalled, and everything was reconnected.

Moving Forward

We are now working on putting together the electronics box, which will contain ADC, and raspberry pi's. The frequency counters will be mounted on the front of the box.

Once complete, it will be installed for use in FOL.

ELOG V3.1.3-