Entry  Wed Jul 16 01:18:02 2014, rana, Summary, LSC, Python Wavelet peak finding for dramatic ALS - Red Resonance finding speedup findRedResonance.pdf
    Reply  Wed Jul 16 01:35:16 2014, Koji, Summary, LSC, Python Wavelet peak finding for dramatic ALS - Red Resonance finding speedup 
       Reply  Mon Jul 21 17:32:43 2014, ericq, Summary, LSC, Arm losses 
Message ID: 10211     Entry time: Wed Jul 16 01:35:16 2014     In reply to: 10209     Reply to this: 10248
Author: Koji 
Type: Summary 
Category: LSC 
Subject: Python Wavelet peak finding for dramatic ALS - Red Resonance finding speedup 

From the last plot:

- Subtracting the offset of 0.0095, the modulation depth were estimated to be 0.20 for 11MHz, 0.25 for 55MHz

- Carrier TEM00 1.0, 1st order 0.01, 2nd order 0.05, 3rd order 0.002, 4th order 0.004

==> mode matching ~93%, dominat higher order is the 2nd order (5%).

Eric: now we have the number for the mode matching. How much did the cavity round-trip loss be using this number?

ELOG V3.1.3-