Message ID: 10147     Entry time: Mon Jul 7 22:07:29 2014     In reply to: 10145
Author: rana 
Type: Update 
Category: Electronics 
Subject: RF cables rerouted 

This Red Ladder movement also probably included some bumping of the MC2 stack (be more careful when working in the lab). Around 5 PM today, the MC lost lock.

When I came in later, I found that the MC was flashing the TEM17,9 mode and had been misaligned like this for ~1 hour. I looked through the MC SUS sensor trends and moved MC2 back to its old position to get the locking back. I disabled the WFS, tweaked the TEM00 mode alignment using only MC2 sliders, and then re-enabled WFS. Its been OK for the past 5 hours.