Entry  Thu Jun 19 14:41:00 2014, Manasa, Update, PSL, ISS disabled 
    Reply  Tue Jun 24 13:02:54 2014, Manasa, Update, PSL, Ringdown PD installed PSL_ringdown.png
       Reply  Mon Jun 30 22:40:21 2014, Manasa, Update, PSL, AOM ringdown AOM_ringSample1.pdf
Message ID: 10115     Entry time: Mon Jun 30 22:40:21 2014     In reply to: 10091
Author: Manasa 
Type: Update 
Category: PSL 
Subject: AOM ringdown 

Quote:

Quote:

I would like to measure the switching time of the AOM. So I have disconnected the modulation input to the AOM that comes from the ISS. I have also turned OFF the SR560's and the AWG that belong to ISS. 

Pics and cable connections of the state in which the ISS setup was left at, will be updated soon.

I installed a fast PDA10CF along the path of a leaking beam from one of the steering mirrors that direct the main beam to the PMC. This beam was dumped to a razor blade. I removed the razor blade and installed a Y1 to steer this beam through a lens on the PD.

Pics of the layout post-installation will be updated.

Also, I tested the AOM by giving it 0-1V modulation input from the AWG. This has been disconnected after the test. So everything should be as it was pre-testing.

Edit/manasa/ Data has not been fit correctly in here. A proper fit will follow this elog.

Proper fits and numbers are here :elog

Earlier last week I had tried to measure the AOM ringdown and concluded I could not make one.

I was proved wrong and I was able to make a measurement. I am still not sure why I was not able to make the measurement earlier with the very same settings and configuration.

What I did:

I gave the AOM a 0-1V modulation input using the signal generator (50 ohm feedthrough bnc was used to impedance match the AOM driver's modulation input). For the measurement here I used a 1Hz square wave. I used a 300MHz oscilloscope to look at the falling edge of the ringdown PD output installed.

I recorded a few ringdown samples. To get a quick number, I fit one such sample to find the AOM switching time as 1.48us (Plot attached). 

Attachment 1: AOM_ringSample1.pdf  653 kB  | Hide | Hide all
AOM_ringSample1.pdf