40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Fri Jun 20 11:41:18 2014, Nichin, Update, Electronics, Transimpedence measurement-BBPD BBPD_readings_06-19-2014.zipBBPD_transimpedence_19thJune2014.pdf
    Reply  Fri Jun 20 19:09:23 2014, Koji, Update, Electronics, Transimpedence measurement-BBPD 
       Reply  Mon Jun 23 21:16:14 2014, Nichin, Update, Electronics, Transimpedence measurement-BBPD BBPD_transimpedence_06-23-2014.pdfBBPD_readings_06-23-2014.zip
Message ID: 10085     Entry time: Fri Jun 20 19:09:23 2014     In reply to: 10079     Reply to this: 10089
Author: Koji 
Type: Update 
Category: Electronics 
Subject: Transimpedence measurement-BBPD 

Oh, nice! This must be a new technique to have a higher transimpedance by breaking the PD.

Now both BBPDs are showing abnormally high impedance.
(Remember, you have not revised your
previous entry after my pointing out you have a bug in the code.)

You should break down the measurement into each raw numbers for validation.
And if this high impedance is still true, you should point out what is causing of this anomaly.

ELOG V3.1.3-